Interference in blood coagulation tests on lipemic plasma. Correction using n-hexane clearing复制

用户-LQudvERqMP4 2个月前 73 10 已关闭

DOI:复制

文献链接:复制

其他信息:

M Arambarri, A Oriol, JM Sancho, FJ Roncalés…
Sangre, 1998
europepmc.org
… To eliminate the bias in the plasmatic coagulation test caused by lipemic samples by means of with addition of n-hexane in the ACL-3000 analyzer (Instrumentation Laboratory). … The turbidity caused by lipemia was simulated by addition of increasing quantities of Intralipid 20% (interferent) to samples from patients with normal pathological coagulation tests (with oral anticoagulant therapy). We compared the results of PT (prothrombin time), fibrinogen (derived) and APTT (activated partial thromboplastin time) before and after the …